The
ECD detector consists of a sealed stainless steel cylinder containing radioactive
Nickel-63.
The Nickel-63
emits beta particles ( electrons ) which collide with the carrier gas molecules,
ionizing them in the process. This forms a stable cloud of free electrons
in the ECD cell. When electro-negative compounds ( especially chlorinated,
fluorinated or brominated molecules ) such as carbon tetrachloride, bromoform,
PCBs and pesticides such as DDT enter the cell, they immediately combine
with some of the free electrons, temporarily reducing the number remaining
in the electron cloud. The detector electronics which maintain a
constant current ( of about 1 nanoampere ) through the electron cloud,
are forced to pulse at a faster rate to compensate for the decreased number
of free electrons. The pulse rate is converted to an analog output
which is connected to the data system. The SRI ECD detector can be operated
with either Nitrogen or Argon/5%Methane ( P5 ) makeup gas, and Nitrogen,
P5, or helium carrier as long as the helium flow is less than 10 ml/min.
The ECD offers extreme sensitivity ( parts per trillion for SF6 ). Because
it contains 5 millicuries of Nickel-63, the ECD is covered by a "General
License" requiring a periodic "wipe test" and the filing of a form with
your state's Department of Health. In most states no annual fee is
required. The ECD may be thermostatted from ambient to 375ºC
8690-0020 ECD detector ( Nickel-63
) $5,495.00 |
|